Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
195
Front-Panel Key and SCPI Command
Reference
4.4.4.1.4 Scale Coupling
Allows you to toggle the scale coupling function between On and Off.
Mode: MRECEIVE
Key Path:
Span X Scale
Remote Command: :DISPlay:AMDepth[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]:COUPle
0|1|OFF|ON
:DISPlay:AMDepth[1]|2:WINDow[1]|2|3|4:TRACe:X[:SCALe]:COUPle
?
Preset: ON
State Saved: Saved in instrument state.
Range: Off|On
Dependencies/Couplings: See Restriction and Notes
Notes: Upon pressing the Restart front-panel key or Restart softkey under the Meas Control
menu, the scale coupling function automatically determines the scale per division and
reference values based on the measurement results if this parameter is set to On.
When you set a value to either Scale/Div or Ref Value manually, Scale Coupling
automatically changes to Off.
Example: :DISP:AMD:WIND:TRAC:X:COUP OFF
:DISP:AMD:WIND:TRAC:X:COUP?