EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>E4445A

Agilent Technologies E4445A User Manual

Agilent Technologies E4445A
434 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #187 background imageLoading...
Page #187 background image
Chapter 3 187
Instrument Functions: M - O
Marker ->
Instrument Functions: M - O
3.3.5 Mkr->Span
Sets the start and stop frequencies to the values of the delta markers. The marker is then
set to normal at the center frequency. Only available in
Delta, Span Pair, and Delta Pair
modes, this function is not available if the marker is off, or in Normal mode or when the
span is
Zero Span.
Key Path:
Marker ->
Remote Command:
:CALCulate:MARKer[1]|2|3|4[:SET]:DELTa:SPAN
Remote Command Notes: Select the delta marker mode with
:CALCulate:MARKer[1]|2|3|4:MODE DELTa.
Example: CALC:MARK2:DELT:SPAN sets the start and stop frequencies to the values of
the delta markers of marker 2.
3.3.6 Mkr->CF
Sets the delta marker to the center frequency. Only available in Delta, Span Pair, and Delta
Pair
modes, this function is not available if the marker is off, or in Normal mode or when the
span is
Zero Span.
Key Path:
Marker ->
History: Added with firmware revision A.02.00
Remote Command:
:CALCulate:MARKer[1]|2|3|4[:SET]:DELTa:CENTer
Remote Command Notes: Select the delta marker mode with
:CALCulate:MARKer[1]|2|3|4:MODE DELTa.
Example: CALC:MARK2:DELT:CENT sets the center frequency to the value of the delta
marker center frequency of marker 2.

Table of Contents

Other manuals for Agilent Technologies E4445A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4445A and is the answer not in the manual?

Agilent Technologies E4445A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4445A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals