Chapter 2 55
Instrument Functions: A - L
AMPLITUDE / Y Scale
Instrument Functions: A - L
2.1.12 Ext Amp Gain
Compensates for external gain/loss. The function is similar to the Ref Lvl Offset function,
however the value is considered, along with the maximum mixer level setting, to
determine the attenuation required (10 dB of Attenuation is added for every 10 dB of
External Amp Gain). The gain is subtracted from the amplitude readout so that the
displayed signal level represents the signal level at the input of the external device.
Gains may only be entered with the numeric keypad or programming commands, not the
knob or step keys.
Key Path:
AMPLITUDE / Y Scale
State Saved: Saved in Instrument State, and survives Preset and power cycle
Factory
Default: 0 dB
NOTE Ext Amp Gain is not affected by Factory Preset or power cycle. It can be reset
to the factory default by pressing
System, Restore Sys
Defaults.
Range: 81.90 dB to 81.90 dB
Remote Command:
[:SENSe]:CORRection:OFFSet[:MAGNitude] <relative_power> (in dB)
[:SENSe]:CORRection:OFFSet[:MAGNitude]?
Example: CORR:OFFS:MAGN 7.3 DB
Sets the Ext Amp Gain to 7.3 dB. The only valid suffix is dB. If no suffix is
sent, dB is assumed.

Table of Contents

Other manuals for Agilent Technologies E4445A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4445A and is the answer not in the manual?

Agilent Technologies E4445A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4445A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals