194 Chapter 3
Instrument Functions: M - O
Meas Control
Instrument Functions: M - O
3.5.5 Abort the Sweep or Measurement (Remote Command Only)
Stops any sweep or measurement in progress and resets the sweep or trigger system. A
measurement refers to any of the measurements found in the
MEASURE menu. If the trigger
conditions are met, another sweep is initiated immediately.
If :INITiate:CONTinuous is off (single measure), then :INITiate:IMMediate will
start a new single measurement.
If :INITiate:CONTinuous is on (continuous measure), a new continuous measurement
begins immediately.
The INITiate and/or TRIGger subsystems contain additional related commands.
History: Added in revision A.02.00
Remote Command:
:ABORt
Remote Command Notes: In the continuous measurement mode, the
Restart key is
equivalent to ABORt.
Example: ABOR