Chapter 4 247
Instrument Functions: P - Z
System
Instrument Functions: P - Z
4.9.2.2.3 Factory
Sets the preset type to “Factory.” When you do a preset, all of the factory default
instrument state will be restored. A factory preset switches the analyzer to the Spectrum
Analysis mode and resets the settings of all the modes to the factory defaults (i.e.
Spectrum Analysis Mode with continuous sweep). Refer to
“Preset” on page 211 for more
information.
Key Path: System, Power On/Preset, Preset Type
Annunciation/
Annotation:
Factory
State Saved: Survives
Preset and power cycle, but not saved in Instrument State.
Factory
Default: Mode
Remote Command:
See “Preset Type” on page 245.
Example:
SYST:PRES:TYPE FACT defines the type of preset as the factory preset.
With factory preset selected, use SYST:PRES to do a factory preset.
4.9.2.3 Save User Preset
Saves the current state of the analyzer into the
User Preset state registor for recall when the
instrument Preset Type is set to User and you perform a preset operation. After you save a
state here, you must go to the
Preset Type key and select User in order to have this state used
as the preset state. Refer to
“Preset” on page 211 key description for the default
factory-configuration settings.
Key Path: System, Power On/Preset
Preset
, if the preset type is set to User. See “Save User Preset” on page 214
Annunciation/
Annotation: None
Dependencies/
Couplings: None
SCPI Status Bits/
OPC Dependencies: None
Remote Command:
:SYSTem:PRESet[:USER]:SAVE
Example: SYST:PRES:SAVE
Use SYST:PRES:TYPE USER to set factory preset type to “User.”