66 Chapter 3
GPRS Mobile Test Application
How Do I Change the MS TX Level?
B. Change the MS TX level during a handover.
1. Press the CALL SETUP key.
2. On Control menu, press the Handover Setup (
F5) key.
3. Select and change the MS TX Level.
4. Press the Handover Execute (
F5) key to change the MS TX level, or press the Close
Menu (
F6) key to leave the level unchanged.
5. Use the
More key to check that the MS TX level has been changed on the Call Parms menu
3 of 3.
3
4
1
5