Programming the Status Register System
Status Groups
Data Questionable Calibration Status Group
The Data Questionable Calibration Status Group is used to determine the specific event that set bit 8 in the
Data Questionable Condition Register. This group consists of the Data Questionable Calibration Condition
Register, the Data Questionable Calibration Transition Filters (negative and positive), the Data Questionable
Calibration Event Register, and the Data Questionable Calibration Event Enable Register.