Chapter 1 41
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
The standard event status register is used to determine the specific event that set bit 5 in
the status byte register. To query the standard event status register, send the command
*ESR?. The response is the weighted decimal sum of the bits which are enabled (set to 1).
For example, if bit number 7 and bit number 3 are enabled, the weighted decimal sum of
the 2 bits is 128 plus 8. So the decimal value 136 is returned.
In addition to the standard event status register, the standard event status group also
contains a standard event status enable register. This register lets you choose which bits in
the standard event status register will set the summary bit (bit 5 of the status byte
register) to 1. Send the *ESE <integer> command where <integer> is the sum of the
decimal values of the bits you want to enable. For example, to enable bit 7 and bit 6 so that
whenever either of those bits is set to 1, the standard event status summary bit of the
status byte register will be set to 1, send the command *ESE 192 (128 + 64). The command
*ESE? returns the decimal value of the sum of the bits previously enabled with the *ESE
<integer> command.
The standard event status enable register presets to zeros (0).
0 A 1 in this bit position indicates that all pending operations were
completed following execution of the *OPC command.
1 This bit is for GPIB handshaking to request control. Currently it is set
to 0 because there are no implementations where the spectrum
analyzer controls another instrument.
2 A 1 in this bit position indicates that a query error has occurred.
Query errors have SCPI error numbers from -499 to -400.
3 A 1 in this bit position indicates that a device dependent error has
occurred. Device dependent errors have SCPI error numbers from
-399 to -300 and 1 to 32767.
4 A 1 in this bit position indicates that an execution error has occurred.
Execution errors have SCPI error numbers from -299 to -200.
5 A 1 in this bit position indicates that a command error has occurred.
Command errors have SCPI error numbers from -199 to -100.
6 A 1 in this bit position indicates that the LOCAL key has been
pressed. This is true even if the instrument is in local lockout mode.
7 A 1 in this bit position indicates that the instrument has been turned
off and then on.

Table of Contents

Other manuals for Agilent Technologies EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies EXA N9010A and is the answer not in the manual?

Agilent Technologies EXA N9010A Specifications

General IconGeneral
BrandAgilent Technologies
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals