Chapter 1 49
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
you want to enable.
Questionable Calibration Positive Transition This command determines which bits in the
Questionable Calibration Condition register will set the corresponding bit in the
Questionable Calibration Event register when the condition register bit has a positive
transition (0 to 1). The variable <integer> is the sum of the decimal values of the bits that
you want to enable.
Questionable Calibration Skipped Register
Questionable Calibration Skipped Condition This query returns the decimal value of the
sum of the bits in the Questionable Calibration Skipped Condition register.
NOTE The data in this register is continuously updated and reflects the
current conditions.
Mode All
Remote Command :STATus:QUEStionable:CALibration:NTRansition
<integer>
:STATus:QUEStionable:CALibration:NTRansition?
Example STAT:QUES:CAL:NTR 16384 Alignment is not required.
Preset 0
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All
Remote Command :STATus:QUEStionable:CALibration:PTRansition
<integer>
:STATus:QUEStionable:CALibration:PTRansition?
Example STAT:QUES:CAL:PTR 16384 Alignment is required.
Preset 32767
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All