Chapter 1 55
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
Questionable Calibration Extended Needed Event Query This query returns the decimal
value of the sum of the bits in the Questionable Calibration Extended Needed Event
register.
NOTE The register requires that the associated PTR or NTR filters be set
before a condition register bit can set a bit in the event register. The
data in this register is latched until it is queried. Once queried, the
register is cleared.
Questionable Calibration Extended Needed Negative Transition This command determines
which bits in the Questionable Calibration Extended Needed Condition register will set
the corresponding bit in the Questionable Calibration Extended Needed Event register
when the condition register bit has a negative transition (1 to 0). The variable <integer> is
the sum of the decimal values of the bits that you want to enable.
Questionable Calibration Extended Needed Positive Transition This command determines
which bits in the Questionable Calibration Extended Needed Condition register will set
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All
Remote
Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed[:EVENt]?
Example STAT:QUES:CAL:EXT:NEED?
Preset 0
SCPI Status
Bits/OPC
Dependencies
Sequential command
Mode All
Remote
Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed:NTRansition
<integer>
:STATus:QUEStionable:CALibration:EXTended:NEEDed:NTRansition?
Example STAT:QUES:CAL:EXT:NEED:NTR 2 Align EMI conducted is not required.
Preset 0
Min 0
Max 32767
SCPI Status
Bits/OPC
Dependencies
Sequential command