54 Chapter 1
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
Questionable Calibration Extended Needed Register
Questionable Calibration Extended Needed Condition This query returns the decimal value
of the sum of the bits in the Questionable Calibration Extended Needed Condition register.
NOTE The data in this register is continuously updated and reflects the
current conditions.
Questionable Calibration Extended Needed Enable This command determines which bits in
the Questionable Calibration Extended Needed Condition Register will set bits in the
Questionable Calibration Extended Needed Event register, which also sets bit 14 of the
Questionable Calibration Register. The variable <integer> is the sum of the decimal values
of the bits you want to enable.
Preset 32767
Min 0
Max 32767
SCPI Status
Bits/OPC
Dependencies
Sequential command
Mode All
Remote
Command
:STATus:QUEStionable:CALibration:EXTended:NEEDed:CONDition?
Example STAT:QUES:CAL:EXT:NEED:COND?
Preset 0
SCPI Status
Bits/OPC
Dependencie
s
Sequential command
Mode All
Remote Command :STATus:QUEStionable:CALibration:EXTended:NEEDed:ENABle
<integer>
:STATus:QUEStionable:CALibration:EXTended:NEEDed:ENABle?
Example STAT:QUES:CAL:EXT:NEED:ENAB 2 Can be used to query if an
EMI conducted alignment is needed.
Preset 32767
Min 0
Max 32767