Chapter 1 59
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
variable <integer> is the sum of the decimal values of the bits you want to enable.
Questionable Integrity Event Query This query returns the decimal value of the sum of the
bits in the Questionable Integrity Event register.
NOTE The register requires that the associated PTR or NTR filters be set
before a condition register bit can set a bit in the event register. The
data in this register is latched until it is queried. Once queried, the
register is cleared.
Questionable Integrity Negative Transition This command determines which bits in the
Questionable Integrity Condition register will set the corresponding bit in the
Questionable Integrity Event register when the condition register bit has a negative
transition (1 to 0)
The variable <integer> is the sum of the decimal values of the bits that you want to enable.
Mode All
Remote Command :STATus:QUEStionable:INTegrity:ENABle <integer>
:STATus:QUEStionable:INTegrity:ENABle?
Example STAT:QUES:INT:ENAB 8 Measurement Uncalibrated
Summary will be reported to the Integrity Summary of the
Status Questionable register.
Preset 32767
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All
Remote Command :STATus:QUEStionable:INTegrity[:EVENt]?
Example STAT:QUES:INT?
Preset 0
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All
Remote Command :STATus:QUEStionable:INTegrity:NTRansition
<integer>
:STATus:QUEStionable:INTegrity:NTRansition?