64 Chapter 1
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
Questionable Integrity Uncalibrated Negative Transition This command determines which
bits in the Questionable Integrity Uncalibrated Condition register will set the
corresponding bit in the Questionable Integrity Uncalibrated Event register when the
condition register bit has a negative transition (1 to 0). The variable <integer> is the sum
of the decimal values of the bits that you want to enable.
Questionable Integrity Uncalibrated Positive Transition This command determines which bits
in the Questionable Integrity Uncalibrated Condition register will set the corresponding
bit in the Questionable Integrity Uncalibrated Event register when the condition register
bit has a positive transition (0 to 1). The variable <integer> is the sum of the decimal
values of the bits that you want to enable.
Example STAT:QUES:INT:UNC?
Preset 0
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All
Remote Command :STATus:QUEStionable:INTegrity:UNCalibrated:NTRansition
<integer>
:STATus:QUEStionable:INTegrity:UNCalibrated:NTRansition?
Example STAT:QUES:INT:UNC:NTR 1 Oversweep cleared will be reported to
the Integrity Summary of the Status Questionable register.
Preset 0
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All
Remote Command :STATus:QUEStionable:INTegrity:UNCalibrated:PTRansition
<integer>
:STATus:QUEStionable:INTegrity:UNCalibrated:PTRansition?
Example STAT:QUES:INT:UNC:PTR 1 Oversweep (Meas Uncal) occurred will
be reported to the Integrity Summary of the Status Questionable
register.
Preset 32767
Min 0
Max 32767