Chapter 1 67
Introduction to Programming X-Series Applications
STATus Subsystem (Remote Only Commands)
Questionable Temperature Register
Questionable Temperature Condition This query returns the decimal value of the sum of the
bits in the Questionable Temperature Condition register.
NOTE The data in this register is continuously updated and reflects the
current conditions.
Questionable Temperature Enable This command determines which bits in the Questionable
Temperature Condition Register will set bits in the Questionable Temperature Event
register, which also sets the Temperature Summary bit (bit 4) in the Questionable
Register. The variable <integer> is the sum of the decimal values of the bits you want to
enable.
Questionable Temperature Event Query This query returns the decimal value of the sum of
the bits in the Questionable Temperature Event register.
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All
Remote Command :STATus:QUEStionable:TEMPerature:CONDition?
Example STAT:QUES:TEMP:COND?
Preset 0
SCPI Status Bits/OPC
Dependencies
Sequential command
Mode All
Remote Command :STATus:QUEStionable:TEMPerature:ENABle <integer>
:STATus:QUEStionable:TEMPerature:ENABle?
Example STAT:QUES:TEMP:ENAB 1 Reference Oscillator Oven Cold
will be reported to the Temperature Summary of the Status
Questionable register.
Preset 32767
Min 0
Max 32767
SCPI Status Bits/OPC
Dependencies
Sequential command