98 Chapter 3
Measurement Synchronization with LXI-B
Setting Up and Executing a List Sweep Measurement
16. Enable the MXGâs
LXI Output LAN Event
:LXI:EVENt:OUTPut:LAN:ENABled
âLAN0â,1
This command is sent
to the MXG.
17. Disable LXI Output
LAN Events on the
MXA
:LXI:EVENt:OUTPut:LAN:DISable:ALL This command is sent
to the MXA
18. Put the MXA into
SA mode
:INST:SEL:SA This command is sent
to the MXA.
19. Put the MXA into
single sweep mode.
:INIT:CONT OFF This command is sent
to the MXA.
20. Choose the MXAâs
PTP domain
:LXI:CLOCk:PTP:DOMain 0 This command is sent
to the MXA.
The parameter value
should match the one
used in Task 3.
21. Choose the MXAâs
LXI domain
:LXI:EVENt:DOMain 0 This command is sent
to the MXA.
The parameter value
should match the one
used in Task 4.
22. Move the MXA into
the list sweep
measurement
CONF:LIST This command is sent
to the MXA.
23. Send the MXA a list
of frequencies.
LIST:FREQ 100 MHz, 200 MHz, 300
MHz, 400 MHz, 500 MHz
This command is sent
to the MXA.
Measurements will be
taken at each of these
frequencies.
24. Configure the
MXAâs list sweep
parameters.
LIST:ATT 10dB
LIST:BAND:RES:TYPE FLAT
LIST:BAND:RES 300 kHz
LIST:BAND:VID 3MHz
LIST:SWE:TIME 1e-6 s
LIST:TRIG:DEL 0
LIST:DET RMS
These commands are
sent to the MXA.
25. Make sure the
MXAâs LXI LAN
triggers are disabled
:TRIG:LXI:LAN:DISable:ALL This command is sent
to the MXA.