Agilent 1200 Preparative Pump User Manual 101
Troubleshooting and Diagnostics
6
Leak Test
Maximum Allowed Flow Rate Exceeded for Both Pistons (1
and 2)
Probable cause Suggested actions
1 Wrong solvent. Use Isopropyl alcohol only.
2 Loose or leaky fittings. Ensure all fittings are tight, or exchange
capillary.
3 Contaminated EMPV. Run the EMPV cleaning procedure.
4 Loose pump head screws. Ensure the pump head screws are tight.
5 Large leaks (visible) at the pump inlet/outlet
valves.
Tighten or exchange the pump inlet/outlet
valves.
6 Large leaks (visible) at filters, multi
assembly, capillaries or EMPV.
Ensure the leaky components are installed
tightly. Exchange the component if required. Run
the EMPV cleaning procedure.
7 Leaky pressure sensor. Exchange the multi assembly.
8 Leaking seal or scratched plungers of
channel 1 or channel 2.
Exchange the pump seals in both channels.
Check the plungers for scratches. Exchange if
scratched.

Table of Contents

Other manuals for Agilent Technologies G1314B VWD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1314B VWD and is the answer not in the manual?

Agilent Technologies G1314B VWD Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1314B VWD
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals