Leak Test
Maximum Allowed Flow Rate exceeded for Piston 1
Maximum Allowed Flow Rate Exceeded for Piston 2
Probable cause Suggested actions
1 Externally leaking inlet or outlet valve in
channel 1.
Ensure the valves are installed correctly. Tighten
the valves.
2 Internally leaking inlet valve in channel 1
(back flow).
Clean or replace the inlet valve in channel 1.
3 Internally leaking outlet valve in channel 2
(back flow).
Clean or replace the outlet valve in channel 2.
4 Leaking seal or scratched plunger in channel
1.
Exchange the pump seals in channel 1. Check
the plunger for scratches. Exchange if
scratched.
5 Loose or leaky fitting or capillary or fitting in
channel 1.
Tighten fitting or replace capillary if necessary in
channel 1.
Probable cause Suggested actions
1 Externally leaking inlet or outlet valve in
channel 2.
Ensure the valves are installed correctly. Tighten
the valves.
2 Internally leaking inlet valve in channel 2
(back flow).
Clean or replace the inlet valve in channel 2.
3 Internally leaking outlet valve in channel 1
(back flow).
Clean or replace the outlet valve in channel 1.
4 Leaking seal or scratched plunger in channel
2.
Exchange the pump seals in channel 2. Check
the plunger for scratches. Exchange if
scratched.
5 Loose or leaky fitting or capillary or fitting in
channel 2.
Tighten fitting or replace capillary if necessary in
channel 2.

Table of Contents

Other manuals for Agilent Technologies G1314B VWD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1314B VWD and is the answer not in the manual?

Agilent Technologies G1314B VWD Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1314B VWD
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals