EasyManua.ls Logo

Agilent Technologies U1242A - Page 66

Agilent Technologies U1242A
85 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
52 U1241A/U1242A User’s and Service Guide
5 Performance Tests and Calibration
9 Turn the rotary switch to the next function according to the Test
Function column shown in Table 5- 14. Repeat steps 3 to 8 for each
adjustment point shown in the calibration adjustment, see Table 5- 15.
10 Verify the adjustments using the Performance Verification Tests.
Table 5 - 15 Calibration Adjustments
Step Test Function Cal Range Input Reference Value Cal Item
U1241A U1242A
1 Turn the rotary switch to V
position
Short Dual banana plug with
copper wires short
between two terminals
SHrt
1000 mV 1 V 1000.0 mV
10 V 10 V 10.000 V
100 V 100 V 100.00 V
1000 V 1000 V 1000.0 V
2 Press to go to V 1000 mV 30 mV, 70 Hz 30.0 mV
function 1000 mV, 70 Hz 1000.0 mV
1000 mV, 1 kHz 1000.0 mV
10 V 1 V, 70 Hz 1.000 V
10 V, 70 Hz 10.000 V
10 V, 1 kHz 10.000 V
100 V 10 V, 70 Hz 10.00 V
100 V, 70 Hz 100.00 V
100 V, 1 kHz 100.00 V
1000 V 100 V, 70 Hz 100.0 V
1000 V, 70 Hz 1000.0 V
1000V, 1 kHz 1000.0 V

Table of Contents

Other manuals for Agilent Technologies U1242A

Related product manuals