50 U1241B/U1242B User’s and Service Guide
5 Performance Tests and Calibration
Valid Adjustment Input Values
Adjustment can be accomplished using the following input values below.
Table 5 -3 Valid adjustment input values
Function Range Valid Input Reference Values
1000 mV, 10 V, 100 V, 1000 V
0.9 to 1.1 x Full Scale
1000 mV, 10 V, 100 V, 1000 V
0.9 to 1.1 x Full Scale
1 V
0.9 to 1.1 x Full Scale
Ω
1000 W, 10 kW, 100 kW, 1000 kW, 10 MW
0.9 to 1.1 x Full Scale
1000 nF, 10 μF, 100 μF, 1000 μF, 10 mF
0.9 to 1.1 x Full Scale
1000 μA, 10000 μA
0.9 to 1.1 x Full Scale
100 mA, 1000 mA
0.9 to 1.1 x Full Scale
10 A
0.9 to 1.1 x Full Scale
T1
0 °C Ensure that 0 °C with ambient
compensation is provided
DCmV
(T1)
100 mV
0.9 to 1.1 x Full Scale
The minimum AC current output of Fluke 5520A calibrator is 29 μA. Ensure that at least 50
μA is set at the calibrator source of AC μA.