50 U1241B/U1242B User’s and Service Guide
5 Performance Tests and Calibration
Valid Adjustment Input Values
Adjustment can be accomplished using the following input values below.
Table 5 -3 Valid adjustment input values
Function Range Valid Input Reference Values
1000 mV, 10 V, 100 V, 1000 V
0.9 to 1.1 x Full Scale
1000 mV, 10 V, 100 V, 1000 V
0.9 to 1.1 x Full Scale
1 V
0.9 to 1.1 x Full Scale
Ω
1000 W, 10 kW, 100 kW, 1000 kW, 10 MW
0.9 to 1.1 x Full Scale
1000 nF, 10 μF, 100 μF, 1000 μF, 10 mF
0.9 to 1.1 x Full Scale
1000 μA, 10000 μA
0.9 to 1.1 x Full Scale
100 mA, 1000 mA
0.9 to 1.1 x Full Scale
10 A
0.9 to 1.1 x Full Scale
T1
0 °C Ensure that 0 °C with ambient
compensation is provided
DCmV
(T1)
100 mV
0.9 to 1.1 x Full Scale
V
CAUTION
The minimum AC current output of Fluke 5520A calibrator is 29 μA. Ensure that at least 50
μA is set at the calibrator source of AC μA.

Table of Contents

Other manuals for Agilent Technologies U1242B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies U1242B and is the answer not in the manual?

Agilent Technologies U1242B Specifications

General IconGeneral
BrandAgilent Technologies
ModelU1242B
CategoryMultimeter
LanguageEnglish

Related product manuals