6 Specifications and Characteristics
94 U3402A User’s and Service Guide
DC Current
[1]
Typical at full scale reading and voltage across the input terminals.
[2]
Use Rel operation.
Ta b l e 6- 2 DCI resolution, full scale reading, and accuracy [± (% of reading + count)]
Rate Range Resolution Maximum
reading
Accuracy
(One year;
23ºC ± 5ºC)
Burden voltage
[1]
and shunt resistor
Slow
12.0000 mA
120.000 mA
1200.00 mA
12.0000 A
0.1 µA
1 µA
10 µA
100 µA
11.9999
119.9999
1199.99
11.9999
0.05% + 15
[2]
0.05% + 5
0.2% + 5
0.2% + 5
<0.15 V/10
W
<1.5 V/10
W
<0.3 V/0.1
W
<0.6 V/0.01
W
Medium
40.000 mA
120.00 mA
1200.0 mA
12.000 A
1 µA
10 µA
100 µA
1 mA
39.999
119.99
1199.9
11.999
0.1% + 6
0.1% + 3
0.2% + 3
0.2% + 3
<0.5 V/10
W
<1.5 V/10
W
<0.3 V/0.1
W
<0.6 V/0.01
W
Fast
40.00 mA
120.0 mA
1200 mA
12.00 A
10 µA
100 µA
1 mA
10 mA
39.99
119.9
1199
11.99
0.1% + 2
0.1% + 2
0.2% + 2
0.2% + 2
<0.5 V/10
W
<1.5 V/10
W
<0.3 V/0.1
W
<0.6 V/0.01
W
U3402-90001.book Page 94 Friday, July 24, 2009 4:04 PM