19.DEVICESTATUSALERT
IfthebatteriesarerunninglowinaCallPoint,Pullcord,DoorMonitororPendantthiswillbedisplayedon
thepanelasshownbelow;
ThescreenbanneriswhiteforaDeviceStatusalert.
TouchthebannertoviewdetailsoftheDeviceStatusalert.
Thiswillidentifyex
actlywhichdevicehasalowbattery.
ADeviceStatusalertcannotbeAcceptedfromthescreen.
TOCANCELTHEALERT
IftheunithasadisplayitwillfirstshowCALLING‐CANCELthenACKNOWLEDGEDwhenthecallhas
beenreceivedandcancelledofftheNursecallpanel.
Ifthereisanothercallwaiting(withinthesameZone)thedetailswillbedisplayed,ifmorethanonecallis
waitingtheonewithhighestprioritywillbedisplayed.
Alwaysreplacethebatterieswithin72hoursofalowbatterywarning.
IfthewarningchangestoCRITICALreplacethebatteriesIMMEDIATELY.
YoumustvisittheCallPoint
andpresstheLIGHTGREY
buttontocancelthealert