17 1800223 E
CHAPTER 4
Operating Instructions
Pre-Use Check-out Procedure (each time before
starting)
Each time before starting operation of the generator,
perform the following inspections. Any failure of these
inspections must be addressed and corrected before
attempting to start the generator.
1. Inspect power cord for degradation, frays,
prong damage, etc. Replace the power cord if
necessary. Use only the ATI supplied power
cord (See Appendix B Spare Parts for
replacement).
2. Inspect liquid aerosol reagent container label
for proper identification, expiration date, etc.
Also inspect liquid aerosol reagent for
cleanliness and transparency.
3. Place the generator in the location where it will
be used. Position the generator so controls are
readily accessible.
4. Verify that aerosol tank output is not blocked
Caution
Do not lift the 6D unit using the sanitary flange or cord
wrap bracket. Damage to the unit may occur.
Caution
Do not obstruct ventilation openings on the front, sides,
or rear panels.
Do not operate unit with less than 6 inches (15cm)
clearance from nearby objects on all four sides.
Doing so may cause the unit to overheat