20 1800223 E
Shutdown and Transportation Procedures
1. Turn the Power Switch to OFF.
2. Disconnect the power cord and any
connections to the generator tank output.
3. ATI recommends emptying the generator tank
or plugging the generator tank output with a
leak-tight plug to prevent oil loss during moving
or transporting.
4. Multiple aerosol generators may be stacked up
to three units high using the foam shipping top
inserts.
Caution
The foam stacking packaging inserts are intended to
allow the units to nest into one another and provide
vertical support only. They are not intended for use
during transportation of units on a cart or vehicle without
the use of additional support or tie-downs.
They are appropriate for use independent from additional
support or retention only in a stationary storage situation
Do not stack generator units more than three units high.
Changing to Different Liquid Aerosol Reagents
Before changing to a different liquid aerosol reagent,
perform the following purging procedure.
1. Drain all previous liquid aerosol reagents from
reservoir.
2. Tilt the unit once each direction side-to-side and
also once front and back to free any trapped oil.
3. Replace drain plug and tighten.
4. Pour denatured alcohol into the generator tank
and tilt the unit to rinse the tank.
5. Remove the drain plug again and tilt the unit each
direction side-to-side and also once front and back
to free any trapped alcohol.
6. Replace drain plug and tighten. The 6D is ready
for use with a different liquid aerosol reagent.