30 1800223 E
APPENDIX D
Troubleshooting
Note
If any remedy listed is not effective, contact ATI for repair
service.
No Aerosol
Output
Liquid aerosol reagent reservoir is
empty
Fill the reservoir with
the appropriate liquid
aerosol reagent
Main Power Switch in OFF
position
Turn Main Power
Switch to the ON
position
Internal filter is clogged
Return the unit for
service
reach operating
pressure
Low air pressure
Increase air pressure
using Jet Pressure
knob
Not plugged to appropriate
voltage source
Verify the unit is
connected to an output
with the correct voltage
Internal air tube is blocked, kinked
or leaking
Return the unit for
service
Internal air filter is clogged
Return the unit for
service
Pressure gauge is defective
Return pressure gauge
for replacement
Desired Aerosol
Output
Low air pressure
Increase air pressure
using Jet Pressure
knob
Unit set to 2 Jets
Turn Jet Select valve to
6 Jets
Laskin Nozzle is blocked or
partially blocked
Return unit for service