31 1800223 E
Symptom
Probable Cause
Remedy
Fill cap is loose or is leaking
Make sure fill cap is
hand tight or does not
leak. Replace if
necessary.
Internal air tube is blocked, kinked
or leaking
Return the unit for
service
Internal air filter is clogged
Return the unit for
service
Unit does not
power on
Not plugged to appropriate
voltage source
Verify the unit is
connected to an output
with the correct voltage
Main Power Switch in OFF
position
Turn Main Power
Switch to the ON
position
Electrical component defective
Return the unit for
service
Unit makes loud
pulsing noise
Excessive air pressure
Pressure relief valve
opens at approximately
23psi. Reduce pressure
below this threshold
Aerosol output port is blocked Remove block
Internal air tube is blocked, kinked
or leaking
Return the unit for
service