PRODUCT DESCRIPTION
4. Technical Data
4.1 Image plate scanners
Power Unit Electrical Details J1100 J1200
Voltage VAC 100 - 240 100-240
Maximum fluctuation in rated voltage % +/- 10
Frequency Hz 50/60
Maximum Current A 0.35-0.15 0.46-0.19
Electrical Data ScanX J1100 J1200
Voltage VDC 24
Maximum fluctuation in rated current % +/- 10
Maximum Current A 1.25 1.65
Power consumption W <30 <39.6
Duty Cycle
S2 min 25
S6 % 25
ScanX Physical Characteristics
Dimensions (W x H x D) in 8.7 X 12.6 X 10.6
cm 22 X 32 X 27
Weight (approximate) lbs 17.6
kg 8.0
Noise Levels
Ready db(A) <40
Scanning db(A) <60
General Values
Medical appliance category 1
Laser protection class IEC 60825-1 Ed. 3 (2014))
Wavelength 635 nm10 mW EN 60825-1:2006-04
Heat radiation W <40
Pixel size can be selected from range µm 21 to 50
Maximum Resolution (depending on plate) hori-
zontal/vertical
Lp/mm 13/20
Conformity Acceptance Procedure
Medical device 2007/47/EC
Barrier envelope Medical device classification I
Plate Medical device classification IIa
EMC
Conducted HF - disturbance V
1
according to IEC 61000-4-6
V
eff
3
Radiated HF - disturbance V
1
according to IEC 61000-4-3
V
eff
4