EasyManua.ls Logo

Air Techniques ScanX Touch - Plate Cleaning, Disinfection, Disposal; Preparing Plates for Patient Use

Air Techniques ScanX Touch
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Air Techniques, Inc.
Page
36
Operation
11.7 Preparing intraoral plates for patient use.
Insert the erased plate into the barrier envelope so the printed side of the PSP is visible through the transparent
side of the envelope. Peel off the adhesive strip and seal the envelope.
View A.
Insert the Plate.
View B.
Peel off the adhesive strip.
View C.
Seal the Barrier Envelope.
11.4 Cleaning phosphor storage plates.
For the best images, plates should be handled carefully and kept
clean. Use specially formulated PSP Cleaning Wipes (P/N B8910)
to clean all plates. These single-use, extra soft, 100% polyester
fabric wipes will not scratch or damage while safely removing dust,
hair, dirt and smudges from the imaging surface. Use one wipe and
clean plates as follows:
1. Gently wipe the PSP Cleaning Wipe over the dry
Plate surface. Wipe back and forth and then in a
circular motion.
2. Allow the plate surface to air dry. Make sure that the
PSP is completely dry before re-using.
Plate must not be put in an autoclave.
Always use standard infection control procedures when
handling devices that contact the patient.
11.5 Disinfecting the phosphor storage plates.
There is no reason to routinely disinfect the PSPs unless contamination is suspected. If a PSP has touched
a contaminated surface, it may be immersed briefly in a cold sterilant (such as a 2% Gluteraldehyde
solution) according to sterilant manufacturers directions. Do not immerse the plate if there is any
evidence of deep scratches in the surface of the plate or nicks in the edges of the plate. After disinfection,
clean and dry the plate using the instructions above.
11.6 Disposal of phosphor storage plates.
Consult with your federal, national, state and local government, for rules and regulations on disposal
of Phosphor Storage Plates.
PSP Cleaning
Sealing the Plate in the Barrier Envelope
The Barrier Envelope must be used only once and
disposed of properly in accordance with local code.