Troubleshooting
Problem Probable cause Solution
X-ray image split vertically
into two halves
Dirt in the laser slit (e.g. hair, dust Clean the laser slit.
X-ray image with small bright
spots or clouding
Micro scratches on the PSP Replace the PSP.
Lamination of the PSP
detaches at the edge
Wrong retainer system used Only use original PSP and lm retainer systems.
PSP handled incorrectly Use the PSP correctly.
Comply with the operating instructions of the
PSP and lm retainer systems.
15.2 Software error.
Problem Probable cause Solution
Too much ambient light Unit is exposed to too much light Darken the room.
Turn the unit such that no light can directly enter
into the entry slot.
Overtemperature Laser or erasure unit too hot Switch the unit o󰀨 and allow it to cool.
Erasure unit fault LED defective Contact technician.