8 Technical Reference 8.4 Interface Description
Signal processing in STARTING, RUNNING and
CLEANING:
If the circuit is broken the separator should
be stopped automatically by NORMAL STOP.
This is to minimise the risk of having access to
moving parts.
Fluid connections
Complementary information is given in the
document Connection List.
201 Inlet
Processing in STAND S TILL:
Shall be closed.
Processing in STARTING:
Should be closed. Bowl will be open and empty
or closed and filled depending on if start is done
from STAND STILL or STOPPING.
Processing in RUNNING:
Could be closed or open.
Processing in CLEANING:
A sequence of cleaning liquids should be fed
to the separator. The flow rate should be as
high as possible and preferably not less than
the production flow rate.
Processing in NORMAL STOP or EMERGENCY
STOP:
Could be closed or open but the bowl shou
ld be
filled unless the stop is initiated in
STARTING.
Processing in SAFE T Y STOP:
Could be closed or open but the bowl shall be
filled unless the stop is initiated in STARTING.
206 Inlet for conditioning and displacement
liquid
According to process
220, 221 and 222 Outlets
Processing in STAND S
TILL:
Could be closed or open.
Processing in other modes:
Shall be open.