Electromagnetic compatibility| AM-5000 User's Manual
Electromagnetic Compatibility
Without approval from Alltion (Guangxi) Instrument Co., Ltd., it may result in the electromagnetic
compatibility of the device or other equipment if it is not authorized to change or refit the device.
The design and test of AM-5000 surgical microscope comply with relevant operating instructions
on electromagnetic compatibility.
The equipment or system shall not be adjacent to or stacked together with other equipment. If it
is required, observe and verify whether it can operate correctly in such a configuration.
Requirements on Wire Installation
Name of Wire
Type
Length (m)
The adapter power cord
Non-shielded parallel wire
3m
Key Components for Electromagnetic Compatibility
The electromagnetic compatibility key components of the product include the power adapter
and dimming circuit board, It will cause significant decreasingly in electromagnetic
compatibility transmission and immunity performance to use or replace with the accessories
with non-matched design,
Do not replace the components without authorization.
Guidance and Manufacturers Declarationelectromagnetic Emission
The AM-5000 surgical microscope is intended for use in the electromagnetic environment
specified below. The customer or the user of the AM-5000 surgical microscope should assure
that it is used in such an environment.
Emissions test
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR11
Group 1
AM-5000 surgical microscope uses RF energy only for
its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR11
Class A
AM-5000 surgical microscope is suitable for use in a
ll establishments, including domestic establishments an
d those directly connected to the public low-voltage po
wer supply network that supplies buildings used for do
mestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations
/flicker emissions
IEC 61000-3-3
Complies