User Interface | 4-3
Zero Key
Use the Zero Key to define on-demand zero settings.
Hold Outputs
Enable
Select this function to hold the output at the last valid reading during a
zero.
Disable
Select this function to allow the output to change during a zero.
Adjust O set
Enable
Select this function to adjust the offset at the end of a zero cycle.
Disable
Select this function if you do not want to adjust the offset.
Zero
This function initiates a zero cycle.
Abort
This function aborts a zero/verify cycle and returns the analyzer to normal
operation.
Figure 4-3.
Main menu for the Zero
key.
ZERO
Hold Outputs
Enable
Disable
Adjust Offset
Zero
Enable
Disable
Abort