Serial Communications | 5-5
Write Data (H)
The data field of this frame contains the handle for a system variable. See
Figures 5-7a and 5-7b for a list of defined handles for system variables.
Example:
Master: >34H05123.40C[CR]
Slave: A[CR]
Write Cell EEPROM (L)
Copy cell serial number and cell calibration coefficients into cell EEPROM.
Example:
Master: >34LB3[CR]
Slave: A[CR]
Quit (Q)
Zero cycle is terminated immediately and normal operation resumed .
Example:
Master: >34QB8[CR]
Slave: A[CR]
Zero (Z)
Initiates a zero cycle.
Example:
Master: >34Z??[CR] (Note the use of wildcard in place of checksum)
Slave: A[CR]