7-8 | 3050 AM Moisture Analyzer
Installing the new dryer
Minimize the time that the dryer is exposed to the atmosphere after
removing the protective caps. AMETEK recommends that the dryer
be exposed for less than a minute.
1. Insert the replacement dryer into the dryer port on the analyzer. Use
the new VCR gaskets provided with the dryer in each of the fittings.
2. Tighten the VCR fittings until they are finger tight.
3 Install the dryer mounting screws to secure the dryer to the bracket.
4. Tighten both of the VCR fittings so that the nuts are 1/8
th
of a turn past
finger tight.
5. Place the dryer compartment cover onto the rear of the analyzer and
secure with the four captured screws.
6. Enter the production code for the new dryer.
!
CAUTION
Sample
Dryer
From
Sample Dryer
To
Sample Dryer
To Sample Dryer
From Zero Dryer
To Zero Dryer
Sample System
Bracket
From Dryer
To Dryer
Exhaust
Sample In
Zero Module
Figure 7-6b.
Model 3050 AM dryers,