5-4 5000 Multi Point Moisture Analyzer
Table IV: Read and Write System Values
Constants and Polynomial Coefficients
System Value Serial ID Description Path
Moist_Span 12 Cell Calibration Factor
TEST/CONFIGLevel 3 Configure MenuConstants
Moist_Gen 20 Moisture Generator Value
TEST/CONFIGLevel 2 Configure MenuMoist_Gen
Convr_Factor 21 Used to Convert ppmv to ppbv or lb/MMscf
TEST/CONFIGLevel 3 Configure MenuConstants
Dry_CF 44 Dry Cell Frequency (Informational Only)
TEST/CONFIGLevel 2 Configure MenuConstants
Cell_SN 45 Cell Serial Number (Informational Only)
TEST/CONFIGLevel 2 Configure MenuConstants
MoistGen_SN 46 Moisture Generator Serial Number
TEST/CONFIGLevel 2 Configure MenuConstants
(Informational Only)
PolyCoeff_0 S0 Calibration Value for Calculating Moisture
TEST/CONFIGLevel 3 Configure MenuPolyCoef
PolyCoeff_1 S1 Calibration Value for Calculating Moisture
TEST/CONFIGLevel 3 Configure MenuPolyCoef
PolyCoeff_2 S2 Calibration Value for Calculating Moisture
TEST/CONFIGLevel 3 Configure MenuPolyCoef
PolyCoeff_3 S3 Calibration Value for Calculating Moisture
TEST/CONFIGLevel 3 Configure MenuPolyCoef
PolyCoeff_4 S4 Calibration Value for Calculating Moisture
TEST/CONFIGLevel 3 Configure MenuPolyCoef
PolyCoeff_5 S5 Calibration Value for Calculating Moisture
TEST/CONFIGLevel 3 Configure MenuPolyCoef
PolyCoeff_A SA Calibration Value for Calculating Moisture
TEST/CONFIGLevel 3 Configure MenuPolyCoef
PolyCoeff_B SB Calibration Value for Calculating Moisture
TEST/CONFIGLevel 3 Configure MenuPolyCoef
PolyCoeff_C SC Calibration Value for Calculating Moisture
TEST/CONFIGLevel 3 Configure MenuPolyCoef
PolyCoef_D1 SD Offset Value for Calculating Moisture
TEST/CONFIGLevel 3 Configure MenuCoef_D
PolyCoef_D2 SE Offset Value for Calculating Moisture
TEST/CONFIGLevel 3 Configure MenuCoef_D
PolyCoef_D3 SF Offset Value for Calculating Moisture
TEST/CONFIGLevel 3 Configure MenuCoef_D
PolyCoef_D4 SG Offset Value for Calculating Moisture
TEST/CONFIGLevel 3 Configure MenuCoef_D
PolyCoef_D5 SH Used for Special Applications Only
TEST/CONFIGLevel 3 Configure MenuCoef_D

Table of Contents