4-8 | Model 900 ADA / Model 930 Sulfur Pit Analyzers
Navigation Examples
Navigating in RUN mode (example):
To view one of the menus in RUN mode:
1. From RUN mode, enter the command for the screen you wish to
view.
Example:
To view the temperatures of each of the analyzer’s temperature
zones, press F6 8. The Show MAI screen will be displayed, and
will default to the first temperature zone, ‘s/lT1’ (Sample Line).
To view other temperature zones, press the other numerical keys
2..8. You do not have to press F6 8 again.
2. To view a different RUN mode screen, enter the command for that
screen. If the new screen has sub-menus, press one of the numeric
keys to view other data.
Navigating in CFG mode (example):
To change the existing temperature Set Point value for the Analyzer
Oven temperature zone:
1. From RUN mode, press F6 –. The message “PSWD1” appears.
2. Press • • (factory default password) or enter the new password, if
changed. An “*” is displayed for each character entered.
3. Press Ent to view the CFG mode screen. “CFG” will be displayed
on the top line, while the output results from RUN mode will con-
tinue to be displayed on the bottom line. This is the CFG mode
normal display.
4. Press F4 0 to display the TStPt (Temperature Set Point) screen. This
menu defaults to parameter ‘1’ (Sample Line Set Point tempera-
ture). Press ‘4’ to display the Oven Set Point temperature.
5. Press Del to delete the existing value (default = 155.00 °C), and
then enter the new value (for example, ‘153.00’) at the “TStPt4?
prompt.
6. To save the change, press Ent and then Esc to back out of this
menu. At the “Save ConFig?” prompt, press Ent again to confirm
the change.
To discard the changes, press Esc.