EasyManua.ls Logo

Ametek 900

Ametek 900
256 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Controller / User Interface | 4-43
Auto-Zero Interval Timer (AZInt)
The AZInt timer sets the interval (hours) between timed Auto-Zeroes.
The timer can be set from 0–999 hours. Set the interval to ‘0’ to turn the
timer off, and disable a timed start of the Auto-Zero. A timed Auto-Zero
will start only when the analyzer is operating in RUN mode without any
Faults.
The Auto-Zero is included in the Auto-Calibration if the Auto-Zero is en-
abled by setting Timer0 to a non-zero condition. If the Auto-Calibrations
and Auto-Zeroes are scheduled to run at the same time, Auto-Calibration
takes precedence.
Example:
To set the AZInt timer to 4 hours and return to CAL mode normal display,
press:
F3 8 Del 4 Ent Ent
When you enter F3 8, the existing interval (hours) and the time re-
maining until the next Auto-Zero (hours/minutes) are displayed.
NOTE

Table of Contents

Related product manuals