Installation and Start-Up | 3-15
6. Terminate the Sample Line and Vent Line wiring as indicated in the
Sample/Vent Line Wiring drawing (Figure 3-6.1) for GP/Div 2 analyz-
ers or the Disconnect Enclosure Wiring drawing (Figure 3-6.2) for
Zone 1 analyzers.
Connect the ground lead from each line to the ground terminals before
connecting the heater leads.
When terminating the temperature sensor (RTD) leads, either sen-
sor in a line can be used as the Control or Over-Temperature sensor
(RTD).
!