AMETEK CTs NetWave
The following steps are required to setup NetWave Multi Source :
- The NetWave Master (L1) is setup in the menu “Standalone/Master/Slave” to the mode “Master (L1). The
Source is set to function “MultiSource”.
The NetWave works as a Master in parallel
mode and is phase L1. The controller
controls the connected Slave NetWave.
- The NetWave Slave (L2) is setup in the menu “Standalone/Master/Slave” to the mode “Slave (L2)”.
The NetWave works as a Slave in parallel
mode and is phase L2. The controller is
inactive. The NetWave hardware is
controlled by the NetWave Master.
- The NetWave Slave (L3) is setup in the menu “Standalone/Master/Slave” to the mode “Slave (L3)”.
The NetWave works as a Slave in parallel
mode and is phase L2. The controller is
inactive. The NetWave hardware is
controlled by the NetWave Master.
- The Framebus OUT of the NetWave Master (L1) is connected to the Framebus IN of the NetWave
Slave (L2)
- The Framebus OUT of the NetWave Slave (L2) is connected to the Framebus IN of the NetWave Slave
(L3)
- The Framebus IN of the NetWave Slave (L3) muss be terminated with a Framebus
Terminator
- The Signal OUT1 of the NetWave Master (L1) is connected to the Signal IN1 of the
NetWave Master (L1)
- The Signal OUT2 of the NetWave Master (L1) is connected to the Signal IN2 of the
NetWave Slave (L2).
-
- The Signal OUT3 of the NetWave Master (L1) is connected to the Signal IN3 of the
NetWave Slave (L3).
- Each NetWave is supplied.
- The PC computer is only connected to the NetWave Master (L1).
- The EUT is connected for Phase L1 to the NetWave Master (L1), for Phase L2 to the NetWave Slave
(L2) and for Phase L3 to the NetWave Slave (L3).
The N from each NetWave are connected.
The PE from each NetWave are connected.