PARSTAT 4000 Hardware Manual 
 
 
PARSTAT 4000 Hardware Manual       21 
 
Figure 7.  Three-Electrode Connection 
 
c) There are some applications for which a four-terminal (Figure 8) connection 
is required.  In this case, the potential between the reference and sense are 
controlled  and  measured.    This  can  be  a  4-terminal  connection  to  a  low 
impedance battery or capacitor, or the Sense electrode can be an additional 
reference  electrode  (for  instance,  in  controlling  the  potential  across  a 
membrane in a H-cell setup). 
 
Note: Running experiments in this mode requires that the Electrometer 
Mode setting in the software be set to Differential for the correct potential 
to be applied and read. 
 
 
 
Figure 8.  Four-Electrode Connection 
 
 
CAUTION:    Take care that the leads do not accidentally short together.  Because 
the black ground lead is often unused, it tends to be overlooked and 
is more likely to accidentally short together with another lead.     
 
 
6.  The polarity of the potential at the counter electrode will be opposite the “applied 
potential”.  This  is  necessary to establish the correct polarity relationship at the 
working electrode versus the reference electrode. 
 
7.  After all other connections to the PARSTAT 4000 are complete, experiments can 
be set up and performed.