Remote Operation
4-4 TM-F2OP-C1XN-01
To set up remote analog programming:
1. Turn the power supply OFF.
2. Using Table 4-2, set switches SW1-1, SW1-2, SW1-3, and SW1-4
according to the selected programming sources. See
the Notes at the
end of
this procedure for more information about switch settings.
3. Install any J2 connector jumpers as indicated in Table 4-2.
4. Connect the programming source(s) to the mating J2
connector as
sho
wn in Figure 4-1 on page 4–6, observing the correct polarity
for
v
oltage sources. Ensure that the mating connector is plugged securely
into the rear panel connector.
5. Set the programming sources to the desired levels and turn the
power
supply ON. The
REM
LED turns on.
6.
Adjust the external programming source
to change the power
supply’
s output.
Notes:
1. Switches SW1 to SW4 can be set to their defa
ults (all open) unless
o
therwise specified in Table 4-2. Switches SW1-5, SW1-6, SW1-7,
and SW1-8 are not required for
remote programming. They can
remain at the settings you
have selected for your application.
2. Resetting switches SW1-3 or SW
1-4, the programming scale
sele
ction switches, may require that you reca
librate the programming
circuit to
maintain programming accuracy. See Chapter 5.
3. Any of the eight switches on SW1 is OFF (OPEN) when it has
been
flip
ped up to break contact, ON (CLOSED) when flipped down to
close contact.
4. The control circuits have been designed to allow you to set
output
v
oltage and current up to 5% over the model-rated maximum valu
es.
The
power supply will operate within these extended ranges, but
we
cannot guara
ntee full performance to specification.