the required sensors are installed.
Derived Parameters:
depth (from pressure)
salc (salinity from conductivity)
salsv (salinity from sound velocity)
density (requires salinity)
sv (sound velocity, requires salinity)
Note: a parameter will only be output if it is
ALSO included in monitor scans. Refer ‘SET
SCAN’ command, above.
5.4 UV Antifouling
Note that these commands apply only in the instance of a UV system installed on the
instrument.
Command
Description
display uv
Print the current UV settings.
set uv time <ontime> <offtime>
Set the duty cycle of the UV lantern in seconds.
Note: Setting UV time to 0 0 will disable light emission.
5.5 Data Logging
Command
Description
display status
Print the current logging status, including in-water status,
number of samples recorded, log duration, and sample
rate.
set logmode [auto | manual]
By default logmode will be set to “auto” and will
automatically start logging when the water immersion is
detected.
logon
Start logging when log mode is set to manual.
logoff
Stop logging when log mode is set to manual.
set filetype [columns | amlx | all]
Set the file type that data will be logged to. The columns
format has file extension “.aml” while the amlx format has
extension “.amlx”. If ALL is selected, two files will be
created, one in each format. See Appendix B for more
details on these two formats.