Command Directory 3-4 Commands
MG369xC GPIB PM PN: 10370-10374 Rev. F 3-11
AMO
Description: Turns off the internal AM function generator if it is on.
Functional
Category: MODULATION
Related
Commands: AM7, AM8
Front Panel Key: N/A
AMR <value> <unit>
Description: Opens the internal AM rate parameter.
Functional
Category: PARAMETER ENTRY
Value: 0.1 Hz to 1 MHz for sine wave; 0.1 Hz to 100 kHz for square, triangle, and ramp
waveforms
Units: MH, KH, HZ
Example: AM7; AMR 10 KH
Turns on internal linear AM with a 10 kHz rate.
Output Command: OAR (in Hz)
Related
Commands: AM7, AM8
Front Panel Key: Modulation | AM (internal) | Edit Rate
AMS <value> <unit>
Description: Opens the external AM %/V sensitivity parameter.
Functional
Category: PARAMETER ENTRY
Value: 0 %/V to 100 %/V
Units: PCV
Example: AM1; AMS 30 PCV
Turns on external linear AM with a sensitivity of 30 percent/Volt.
Output Command: OAS
Related
Commands: AM1
Front Panel Key: Modulation | AM (external) | Edit Sensitivity

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu MG369 C Series and is the answer not in the manual?

Anritsu MG369 C Series Specifications

General IconGeneral
BrandAnritsu
ModelMG369 C Series
CategoryPortable Generator
LanguageEnglish

Related product manuals