Command Directory 3-4 Commands
MG369xC GPIB PM PN: 10370-10374 Rev. F 3-25
D4 <value> <unit>
Description: Opens the internal pulse delay4 parameter.
Functional
Category: PARAMETER ENTRY
Value: 30 ns to 167 ms at 100 MHz pulse clock rate; 100 ns to 419 ms at 40 MHz pulse clock rate;
300 ns to 1.6 s at 10 MHz pulse clock rate
Units: SEC, MS, US, NS
Example: PMD4; D4 100 US; IP
Turns on internal pulse quadruplet with a 100 s delay before the 4th pulse.
Output Command: OD4 (s)
Related
Commands: IP, PMD(x), PTG(x)
Front Panel Key: Modulation | Pulse (internal) | Wdth/Dly List. . . | Edit Selected
DB
Description: Decibel (dB) data terminator.
Functional
Category: DATA TERMINATOR
Front Panel Key: N/A
DF0
Description: Selects F0-delta frequency sweep mode.
Selects a symmetrical frequency sweep around F0. The width of the sweep is determined
by the Delta Frequency parameter.
Functional
Category: SWEEP
Example: SSP; DF0; AUT
Sets step sweep F0-F with auto triggering.
Related
Commands: F0, DLF, DFF, DFM
Front Panel Key: N/A
DF1
Description: Selects F1-delta frequency sweep mode.
Selects a symmetrical frequency sweep around F1. The width of the sweep is determined
by the Delta Frequency parameter.
Functional
Category: SWEEP
Example: SSP; DF1; AUT
Sets step sweep F1-F with auto triggering.
Related
Commands: F1, DLF, DFF, DFM
Front Panel Key: N/A