Range Errors
A listing of Range Error messages is given in Table 3-3.
3-23
Chapter 3 Functions and Operations
Error Message
Description
RANGE
ERROR:F1 > F2
The start (F1) frequency is greater than the stop (F2) frequency.
RANGE
ERROR:D1 > D2
The start (D1) distance is greater than the stop (D2) distance.
RANGE
ERROR:D2 >
DMax=xx.x ft (m)
The stop distance (D2) exceeds the maximum unaliased range. This
range is determined by the frequency span, number of points,
and relative propagation velocity:
MaximumUnaliased Range
dp V
f
FF
=
´
-
( . )( )( )15 10
8
21
Where: dp is the number of data points (128, 256, or 512)
V
f
is the relative propagation velocity
F2 is the stop frequency in Hz
F1 is the start frequency in Hz
Maximum Unaliased Range is in meters
Refer to Figure 3-14, page 3-24, for a plot of maximum range versus
frequency span.
RANGE ERROR:
TOP<=BOTTOM
The SWR scale parameter top value is less than or equal to its bottom
value.
RANGE ERROR:
TOP>=BOTTOM
The RL scale parameter top value is greater than or equal to its bottom
value.
CAL
INCOMPLETE
A complete open, short, and load calibration must be performed bef-
ore calibration can be turned on.
DIST REQUIRES
F1<F2
Valid distance to fault plots require a non-zero frequency span.
DIST REQUIRES
CAL
Distance-to-fault measurements do not provide usable data with CAL
OFF.
NO STORED
SWEEP AT THIS
LOCATION
Attempting to recall a display from a location that has not been previ
-
ously written to. That is, the location does not contain stored sweep.
USE OPTIONS
MENU TO
SELECT A
PRINTER
Attempting to print a display with no printer selected. Select a printer,
then retry.
DISTANCE AND
CABLE
INSERTION LOSS
MODE ARE
INCOMPATIBLE
DTF measurements only display RL or SWR versus distance.
CANNOT ZERO
NO DETECTOR
INSTALLED
Attempting to perform a Power Monitor zero adjust function with no
RF detector connected to the Site Master.
Table 3-3. Range Error Messages (1 of 2)