EasyManua.ls Logo

AP TC5-2V2SA - Page 30

AP TC5-2V2SA
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
30
TC5-2V2SA.rev.05
TC5-2V2SA
n Press the push-button. The relative humidity set point is displayed,
alternating with the letters "set rH".
n Use the adjustment knob to adjust the set point to the desired value.
3
Adjusting Minimum Speed Compensation
This is the percentage increase in minimum speed for a relative humidity
equal to the humidity set point + 10%. The value ranges from 0 to 100%.
n Set the selection knob to STAGE 1 — HUMIDITY CONTROL. The
current humidity reading is displayed.
n Press the push-button twice. The current minimum speed compensation
is displayed, alternating with the letters "SPd".
n Use the adjustment knob to adjust the minimum speed compensation
to the desired value.
4
Activating / Deactivating Humidity Compensation
n Set the selection knob to STAGE 1 — HUMIDITY CONTROL. The
current humidity reading is displayed.
n Press the push-button three times. The current on/off state of humidity
compensation flashes on the display.
n Use the adjustment knob to adjust the on/off state to the desired value.