EasyManua.ls Logo

Aplisens APR-2000ALW - Page 34

Aplisens APR-2000ALW
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E6 32 DTR.APC.APR.ALW.03(ENG)
5.3.4. APR-2000ALW, Metrological parameters
Accuracy ± 0.075% (FSO) of the calibrated range
± 0.1% (FSO) for range10
Special version ± 0.05% of the calibrated range
Long term stability accuracy / 3 years
or 2 x accuracy / 5 years
HS version accuracy for 6 years
Error due to supply voltage changes ± 0.002%(FSO)/1V
Thermal error ± 0.05%(FSO)/10ºC for range n° 19
± 0.08%(FSO)/10ºC for range n° 10
± 0.25%(FSO) for the whole compensation range
Special version ± 0.03%(FSO)/10ºC for range n° 1…9
± 0.1%(FSO) for the whole compensation range for range n° 1…9
Zero shift error for static pressure* ± 0.06 % (FSO)/10bar (for range 1, 2)
± 0.01 % (FSO)/10bar (for range 3, 4, 5, 6, 7, 9)
± 0.03 % (FSO)/10bar (for range 8)
± 0.02 % (FSO)/10bar (for range n°10)
± 0.01 % (FSO)/10bar (for range n°2, 8 in HS version
Cut-off on radical characteristic curve up to10% of flow.
*) Zeroing in static pressure conditions with zero differential pressure eliminates this error.
5.3.5. APR-2200ALW. Metrological parameters
Accuracy ± 0.1% (FSO) for the APR-2200ALW)
The other parameters as given in the p. 5.3.4.
Sealing effect errors as given in the relevant diaphragm seal Catalogue Card concerning the distance seal.
NOTE: The additional absolute zero error due to ambient temperature can be compensated by configuring
the transmitter, seals and capillaries in accordance with the recommendations on pages II/ 23 and
II/ 24 Product Catalogue.
5.3.6. APR-2000ALW Metrological Parameters according to PED version
APR-2000ALW transmitters which conform to PED are characterised by a measurement range of between -
1bar and 250bar, static pressure of 250bar and the overpressure of PS=275bar.
The permissible operating temperature for option according to:
- H1 Module is between TS min/max: -25ºC and 100ºC;
5.3.7. APR-2000ALW. Pressure Connectors
APR-2000ALW C-type connector to mount together with a valve manifold see Fig.10, available for PED
version.
APR-2000ALW, with single direct diaphragm seal as in the example (Fig.11) or with other diaphragm seals in
accordance with “DIAPHRAGM SEALS” catalogue cards.
5.3.8. APR-2200ALW, Pressure Connectors - diaphragm seals
see “DIAPHRAGM SEALS catalogue cards.

Table of Contents