EasyManua.ls Logo

Aplisens APR-2000ALW - Apc-2000 Alw;L; Apr-2200 Alw;D - Measurement Ranges and Metrological Parameters; Easurement Ranges and Metrological Parameters

Aplisens APR-2000ALW
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E6 34 DTR.APC.APR.ALW.03(ENG)
5.6. APC2000ALW/L. Measurement ranges and metrological parameters
5.6.1. APC2000ALW/L…level transmitter. Measurement ranges.
N
1
2
3
Nominal range (FSO)
0...20m H
2
O
0...10m H
2
O
0...2.5m H
2
O
Minimum set range
2m H
2
O
1m H
2
O
0.5m H
2
O
Overpressure limit
0...200m H
2
O
0...100m H
2
O
0...25m H
2
O
5.6.2. APC2000ALW/L…level transmitter. Permitted environmental conditions
Operating temperature range -40º ÷ 85ºC
Medium temperature range -40º ÷ 40ºC
ETFE, PU+PTFE version -40º ÷ 80ºC
ETFE+PTFE version -40º ÷ 10C
5.6.3. APC–2000ALW/L…level transmitter. Metrological parameters.
Accuracy ± 0.16% (FSO) for the APC-2000ALW nominal range
Long term stability accuracy / 2 years
Error due to supply voltage changes ± 0.002%(FSO)/1V
Thermal error ± 0.1%(FSO)/10ºC
± 0.4%(FSO)/ in the whole compensation range.
Thermal compensation range -25…100ºC
-40…80ºC - special version
5.6.4. APC2000ALW/L. level transmitter. Pressure Connectors
- SG-25 sensor;
- SG-25S sensor;
- SG-25C sensor;
- SG-25S-tytan sensor;
- SG-16 sensor;
5.6.5. APC–2000ALW/L…. pressure transmitter.
Pressure range see p. 5.2.1
Metrological parameters see p. 5.6.3
Pressure connectors see p. 5.2.5
5.7. APR2200ALW/D Measurement ranges and metrological parameters
5.7.1. APR2200ALW/D density transmitter. Measurement ranges
N
Nominal measuring
range (FSO)
Nominal range with reference
to measured density
Minimum set range
Ability to shift the start
of the range
1
-70...0 mbar
0…1000 kg/m
3
100 kg/m
3
0…900 kg/m
3
2
-70...70 mbar
0…2000 kg/m
3
200 kg/m
3
0…1800 kg/m
3
5.7.2. APR2200ALW/D. Permitted environmental conditions
Medium temperature range -40 ÷ 80ºC
The medium must not be allowed to freeze in the immediate vicinity of the transmitter.