E6 42 DTR.APC.APR.ALW.03(ENG)
10.2.5. TRANSMITTER CONFIGURATION WITH USING ITS BUTTONS AND LOCAL MENU
10.2.5.1. Local menu - structure. Local configuration of transmitters.
If the option of local configuration is active, operator can change transmitter set using buttons being below
display. The access to buttons will get after unscrewing the display cover. Then we can also change the display
position (see Fig. 4).
If the option of local configuration is active, operator can change transmitter set using buttons being below
display. To enter changes at the local set mode, press one of the buttons and hold it’s about 4s. If after pressing
the button displays the message ERR_L16, local configuration of the transmitter is switched off. To its switching
on is necessary to use KAP 3 Calibrator or PC (see –> HART command 132, 133).
The buttons are signed with symbols: [↑] [↓] [◙]
After pressing by 4 seconds any of the buttons on the display will appear “EXIT”.
If you will confirm this message by pressing and holding button [◙] by 1 sec, you will leave the local change of
the MENU setting. If you do not confirm, you can move in MENU and change interesting you parameters.
Pressing button [↑] moves up in tree's structure MENU.
Pressing button [↓] moves down in tree's structure MENU.
Pressing [◙] confirms choice and leads change.
No action in the menu for more than 2 minutes will cause automatically exit from the menu and move to the
display of the process variable.
Method of navigate through the commands in the menu structure APC..., APR... shown below.