GeneAmp
®
PCR System 2700
For Amplification of Nucleic Acids
User Guide

Table of Contents

Questions and Answers

  • S
    Stacey Wise DVMAug 22, 2025
    What to do if my Applied Biosystems GeneAmp PCR System 2700 Laboratory Equipment has no screen display and no response when turned on?
    • H
      Hannah CowanAug 22, 2025
      If your Applied Biosystems Laboratory Equipment shows no screen display and doesn't respond when you turn it on, check the following: Ensure the power switch is in the ON position, verify the power cord is properly connected, and inspect the fuses.
  • J
    Judith GillespieAug 25, 2025
    What to do if the printer is not responding to Applied Biosystems GeneAmp PCR System 2700 Laboratory Equipment?
    • A
      Arthur LewisAug 25, 2025
      If the printer is not responding to your Applied Biosystems Laboratory Equipment, check the printer connections and the power switch.
  • A
    ashley93Aug 26, 2025
    What to do if control panel is not responding on Applied Biosystems Laboratory Equipment?
    • M
      Michael MoralesAug 26, 2025
      If the control panel on your Applied Biosystems Laboratory Equipment is not responding, check for any keys that may be stuck in the down position and free them.
  • M
    michelle11Aug 27, 2025
    Why does my Applied Biosystems GeneAmp PCR System 2700 Laboratory Equipment printer fail?
    • R
      ramosdawnAug 28, 2025
      Printer failures with Applied Biosystems Laboratory Equipment can occur due to an incorrect printer configuration or cable. Verify the printer settings are set to baud rate = 9600, no parity, one stop bit, and eight data bits. You can also purchase an Applied Biosystems printer cable.
  • E
    edaughertyAug 29, 2025
    Why is my Applied Biosystems GeneAmp PCR System 2700 making too much noise?
    • S
      Sarah LamAug 29, 2025
      If your Applied Biosystems Laboratory Equipment is making too much noise, check for any obstructions in the vents.
  • S
    Sara TorresAug 31, 2025
    Why is the cooling rate too slow in my Applied Biosystems GeneAmp PCR System 2700 Laboratory Equipment?
    • S
      Susan DouglasAug 31, 2025
      A slow cooling rate in your Applied Biosystems Laboratory Equipment can be caused by a warm ambient temperature. Move the instrument to a well-ventilated location with a temperature between 15–30 °C. It also can be caused by Peltier failure.
  • S
    sullivanjessicaSep 2, 2025
    Why is the heating rate too slow on my Applied Biosystems GeneAmp PCR System 2700?
    • D
      Dawn BonillaSep 2, 2025
      A slow heating rate on your Applied Biosystems Laboratory Equipment can be caused by Peltier failure.
  • J
    Jeremy SmithSep 4, 2025
    Why is the cycling time too long for Applied Biosystems Laboratory Equipment?
    • A
      Ashley DavidsonSep 4, 2025
      An extended cycling time in Applied Biosystems Laboratory Equipment can be attributed to Peltier failure.
  • D
    denisemcfarlandSep 6, 2025
    Why can’t my Applied Biosystems Laboratory Equipment reach the high or low temperature range?
    • C
      Christopher HartSep 6, 2025
      If your Applied Biosystems Laboratory Equipment struggles to reach the desired high or low temperature range, this could be due to Peltier failure or an ambient temperature that is too warm.
  • E
    Erika HansenSep 8, 2025
    What to do if Applied Biosystems Laboratory Equipment shows 'Valid range is' error?
    • E
      Elizabeth JenkinsSep 8, 2025
      If your Applied Biosystems Laboratory Equipment displays a 'Valid range is' error, it means you entered a number out of range. Re-enter a value within the specified parameters.

Applied Biosystems GeneAmp PCR System 2700 Specifications

General IconGeneral
Temperature Range4°C to 99°C
Temperature Uniformity±0.5°C
Block Capacity96 x 0.2 mL tubes
DisplayLCD
InterfaceRS-232
Temperature Accuracy±0.25°C (35-99.9°C)
Thermal Cycler TypePeltier-based