95
MEAS:MAXD:K?
Request for final result of Maximum Demand Test.
MEAS:MAXD:K:TEMP?
Request for actual (temporary) value of Maximum Demand Test.
MEAS:MAXD:WIN:BEG?
Request for timestamp of the beginning of Maximum Demand Test window.
MEAS:MAXD:WIN:END?
Request for duration of Maximum Demand Test window.
MEAS:MAXD:RES
Resets Maximum Demand Test.
MEAS:MAXD:RES?
Request for timestamp of last reset of Maximum Demand Test.
SYST:ASS:STAR
Start of Automatic Snapshots Saving Routine with predefined periodicity.
SYST:ASS:STAR?
Request for status of Automatic Snapshots Saving Routine.Answers are: 0 not running, 1 running.
SYST:ASS:STOP
Cease of Automatic Snapshots Saving Routine.
SYST:ASS:STAT?
Request for number of saved snapshots in currently running Snapshots Saving Routine.
SYST:ASS:MAXC {<x>}
Setting the number (x) of snapshots to be made during Automatic Snapshots Saving Routine.
SYST:ASS:MAXC?
Request for the number of snapshots to be made during Automatic Snapshots Saving Routine.
SYST:ASS:PER {<x>}
Setting the period in minutes (x) of snapshots saving occurrence during Automatic Snapshots Saving
Routine. Minimum value is 0.1 (6s) and maximum is 1440 (24h).
SYST:ASS:PER?
Request for the time period of snapshots saving occurrence during Automatic Snapshots Saving
Routine.

Other manuals for Applied Precision WS 2120

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Applied Precision WS 2120 and is the answer not in the manual?

Applied Precision WS 2120 Specifications

General IconGeneral
BrandApplied Precision
ModelWS 2120
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals